Advancing an autoperfusion balloon catheter with the guidewire exiting a distal perfusion sidehole was identified as a cause of coronary artery dissection during angioplasty.
Highlights a previously unrecognized mechanism of coronary artery dissection during angioplasty due to guidewire exit through a distal perfusion sidehole of an auto-perfusion balloon catheter.
Answers are grounded in this paper and cite their sources.
Clinicians should recognize this dissection risk with autoperfusion balloons; extends known mechanisms but leaves open confirmation in prospective studies.
A previously unrecognized cause of coronary artery dissection is reported. A 67-year-old woman underwent angioplasty of the right coronary artery using an autoperfusion balloon catheter. Dissection occurred because the balloon catheter was advanced while the guidewire exited from one of the distal perfusion sideholes.
No takes yet. Share an insight, caveat, or question.
Werns et al. (1994) studied this question. Advancing an autoperfusion balloon catheter with the guidewire exiting a distal perfusion sidehole was identified as a cause of coronary artery dissection during angioplasty.