Key result
Peripheral pulse pressure in the finger and wrist was significantly reduced during head-up-tilt compared to supine posture (P<0.05), whereas carotid pulse pressure was not.
Why the study?
Are peripheral pulse pressure measurements suitable surrogates for carotid pulse pressures during postural stress?
Observational (n=7)
Are peripheral pulse pressure measurements suitable surrogates for carotid pulse pressures during postural stress?
p-value: p=<0.05
Peripheral pulse pressure measurements do not accurately reflect carotid pulse pressures during postural stress, indicating they are not suitable surrogates for central measurements in this context.
Peripheral pulse pressures may not surrogate carotid values during tilt; leaves open need for central measures in postural studies.
OBJECTIVE: Interpretation of baroreflex cardiovascular control requires accurate assessment of pulse pressure (PP) in central arteries under conditions of varying systemic or hydrostatic pressure. The objective of this study was to examine whether changes in PP during postural stress were similar in the peripheral versus carotid arteries. DESIGN: Protocol A: Pulse pressure was measured in both the left (Millar tonometer) and right (Colin Pilot) radial arteries, and in the finger (Finapres) in seated subjects (n = 7) who performed Valsalva's manoeuvre. Protocol B: PP was measured from the carotid (Millar tonometer), and from the finger and wrist kept at the level of the carotid artery, during supine and 60 degrees head-up postures. RESULTS: Protocol A: Pulse pressures during Valsalva's manoeuvre were highly correlated between all devices (r = 0.6-0.8). Protocol B: compared with supine, PP was reduced in both the finger and wrist during head-up-tilt (HUT) (P<0.05), but not in the carotid artery. CONCLUSIONS: During Valsalva manoeuvers the Millar and Colin tonometers similarly tracked PP over a wide range of rapidly changing pressures. This observation provided confidence for the further use of the hand-held device for central measurements during changes in posture. The results from Protocol B indicate that peripheral PP measurements are not suitable surrogates for carotid pulse pressures during HUT.
No takes yet. Share an insight, caveat, or question.
Steinback et al. (2004) reported an observational. Postural stress (head-up-tilt) vs. Supine posture was evaluated on Pulse pressure in peripheral versus carotid arteries (p=<0.05). Peripheral pulse pressure in the finger and wrist was significantly reduced during head-up-tilt compared to supine posture (P<0.05), whereas carotid pulse pressure was not.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: