Key result
Laserballoon energy and application duration increased thermal lesion size in a dose- and time-dependent manner, with steam pops provoked at 12.0 W for >16 seconds.
Why the study?
Although laserballoon-based pulmonary vein isolation is safe and effective, the influence of laser energy titration on lesion formation has never been fully investigated.
Does laser energy titration affect lesion size and steam pop incidence in porcine myocardium?
Does laser energy titration affect lesion size and steam pop incidence in porcine myocardium?
Laserballoon ablation creates lesions in a dose- and time-dependent manner, with high-energy and prolonged irradiation increasing the risk of steam pops.
Answers are grounded in this paper and cite their sources.
May guide laserballoon energy titration to limit steam pops in preclinical models; leaves open optimal parameters and safety in humans.
Laserballoon-based pulmonary vein isolation has proven to be safe and effective. However, the influence of the laser energy titration on the lesion formation has never been fully investigated. The aim of this study was to determine the relationship between the delivered laser energy and lesion size, as well as the incidence of steam pop.The whole porcine heart was excised, and the left ventricular myocardium was separated into four specimens. Myocardial specimens were embedded in a warm mattress to keep the myocardial temperature around 37°C. The laserballoon was located so that the surface of the laserballoon was attached to the myocardium. The laser energy was irradiated against the surface of myocardium at 5.5, 8.5, 10.0, and 12.0 W for 3, 5, 10, and 20 seconds. The depth, surface area, and lesion volume were measured using a digital vernier caliper.At constant laser energy and time, the lesion size increased significantly with the increasing energy (P < 0.001) and application duration (P < 0.001). The steam pop was provoked when a 12.0 W laser energy was applied for longer than 16 seconds, and it occurred in 2 out of 8 lesions.The laserballoon demonstrated the ability to create a lesion formation in a dose- and time-dependent manner. Steam pop could be provoked with high-energy irradiation.
No takes yet. Share an insight, caveat, or question.
Okishige et al. (2020) studied Atrial fibrillation (experimental ablation model) (n=50). Laserballoon energy application vs. Varying energy levels and durations was evaluated on Lesion size (area, depth, and volume) (p=<0.001). Laserballoon energy and application duration increased thermal lesion size in a dose- and time-dependent manner, with steam pops provoked at 12.0 W for >16 seconds.
Synapse has enriched 10 related papers. Consider them for comparative context: