Key result
Resolution limited spatial filtering of ECG gated B-scan echocardiographic images selectively removes high spatial frequency energy, resulting in a significant increase in signal to noise ratio.
Answers are grounded in this paper and cite their sources.
Resolution limited spatial filtering can significantly improve the signal-to-noise ratio in B-scan echocardiographic images by removing high spatial frequency speckle.
Much of the signal energy due to speckle in ultrasound images is shown to be of higher spatial frequency than the intrinsic pulse shape limited resolution in B-scan ultrasound images. A significant increase in signal to noise ratio can therefore be obtained by resolution limited spatial filtering which selectively removes energy of higher spatial frequency than the pulse envelope resolution limit. The concept is illustrated by resolution limited filtering ECG gated B-scan echocardiographic images. Signal to noise improvement is illustrated by comparing time-motion displays generated from both processed and unprocessed images.
No takes yet. Share an insight, caveat, or question.
Parker et al. (1982) studied this question. Resolution limited spatial filtering vs. Unprocessed images was evaluated on Signal to noise ratio. Resolution limited spatial filtering of ECG gated B-scan echocardiographic images selectively removes high spatial frequency energy, resulting in a significant increase in signal to noise ratio.