Key result
Non-invasive cardiac output monitoring shows ~37% error versus thermodilution during hemodynamic variability.
Why the study?
Does non-invasive finger arterial pressure-derived cardiac output measurement agree with thermodilution cardiac output in patients undergoing cardiothoracic surgery across different grades of hemodynamic variability?
Observational (n=27)
No
Does non-invasive finger arterial pressure-derived cardiac output measurement agree with thermodilution cardiac output in patients undergoing cardiothoracic surgery across different grades of hemodynamic variability?
p-value: p=0.035
Hemodynamic variability during surgery significantly increases the error when comparing non-invasive continuous cardiac output monitoring with thermodilution, indicating that future comparison studies must report these variabilities.
Variability increases non-invasive cardiac output error; supports caution in unstable conditions and leaves open method reliability without variability reporting.
Summary While haemodynamic variability interferes with the assumption of constant flow underlying thermodilution cardiac output calculation, variability in (peripheral) arterial vascular physiology may affect pulse contour cardiac output methods. We compared non‐invasive finger arterial pressure‐derived continuous cardiac output measurements (Nexfin ® ) with cardiac output measured using thermodilution during cardiothoracic surgery and determined the impact of cardiovascular variability on either method. We compared cardiac output derived from non‐invasive finger arterial pressure with cardiac output measured by thermodilution at four grades (A–D) of cardiovascular variability. We defined Grade A data as heart rate and mean arterial pressure variability < 5% and the absence of arrhythmias (implying stable flow) and Physiocal ® interval (as measure of variability in finger arterial physiology) > 30 beats. Grade B included all levels of heart rate/mean arterial pressure variability and arrhythmias (Physiocal < 30 excluded). Grade C included all Physiocal intervals (heart rate/mean arterial pressure variability > 5% and arrhythmias excluded). Grade D included all data. Comparison results were quantified as percentage errors. We analysed measurements in 27 patients undergoing coronary artery bypass surgery. Before extracorporeal circulation, the percentage error was 23% (n = 14 patients) in grade A, 28% (n = 20) in grade B, 32% (n = 22) in grade C and 37% (n = 26) in grade D, with a significant increase in variance (p = 0.035). Bias did not differ between grades. After extracorporeal circulation (n = 27), percentage errors became larger, but were not different between grades. Variability during cardiothoracic surgery affected the comparison between thermodilution and non‐invasive finger arterial pressure‐derived cardiac output. When the main sources of variability were included, percentage errors were large. Future cardiac output methodology comparison studies should report haemodynamic variability.
No takes yet. Share an insight, caveat, or question.
Truijen et al. (2018) conducted an observational in Coronary artery bypass surgery (n=27). Non-invasive continuous cardiac output monitoring (Nexfin) vs. Thermodilution cardiac output was evaluated on Percentage error between non-invasive and thermodilution cardiac output across different grades of cardiovascular variability (p=0.035). Non-invasive continuous cardiac output measurement had a 23% percentage error compared to thermodilution under stable conditions, which increased to 37% when hemodynamic variability was included.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: