Key result
Calibration-free cuffless PPG monitoring estimates systolic BP within ~7 mmHg MAE.
Why the study?
Traditional cuff-based sphygmomanometers can be uncomfortable and particularly unsuitable to use during sleep.
Does a calibration-free method combining PPG dynamics and pulse morphology accurately estimate blood pressure compared to reference measurements?
Does a calibration-free method combining PPG dynamics and pulse morphology accurately estimate blood pressure compared to reference measurements?
Combining PPG dynamics with pulse morphology features using neural networks offers a potentially accurate, single-sensor, calibration-free method for continuous noninvasive blood pressure monitoring.
Should not yet change clinical BP monitoring; leaves open larger validation of calibration-free PPG estimation.
Traditional cuff-based sphygmomanometers for measuring blood pressure can be uncomfortable and particularly unsuitable to use during sleep. A proposed alternative method uses dynamic changes in the pulse waveform over short intervals and replaces calibration with information from photoplethysmogram (PPG) morphology to provide a calibration-free approach using a single sensor. Results from 30 patients show a high correlation of 73.64% for systolic blood pressure (SBP) and 77.72% for diastolic blood pressure (DBP) between blood pressure estimated with the PPG morphology features and the calibration method. This suggests that the PPG morphology features could replace the calibration stage for a calibration-free method with similar accuracy. Applying the proposed methodology on 200 patients and testing on 25 new patients resulted in a mean error (ME) of -0.31 mmHg, a standard deviation of error (SDE) of 4.89 mmHg, a mean absolute error (MAE) of 3.32 mmHg for DBP and an ME of -4.02 mmHg, an SDE of 10.40 mmHg, and an MAE of 7.41 mmHg for SBP. These results support the potential for using a PPG signal for calibration-free cuffless blood pressure estimation and improving accuracy by adding information from cardiovascular dynamics to different methods in the cuffless blood pressure monitoring field.
No takes yet. Share an insight, caveat, or question.
Samimi et al. (2023) studied Blood pressure estimation (n=255). PPG morphology-based cuffless blood pressure estimation vs. Invasive and noninvasive reference blood pressure was evaluated on Mean absolute error (MAE) of estimated blood pressure compared to reference measurements. The calibration-free cuffless blood pressure estimation method using PPG cardiovascular dynamics achieved a mean absolute error of 7.41 mmHg for systolic and 3.32 mmHg for diastolic blood pressure in a 25-patient test set.