How can the risk of electromagnetic interference from medical equipment be managed in patients with implanted cardiac devices?
How can the risk of electromagnetic interference from medical equipment be managed in patients with implanted cardiac devices?
This review highlights the importance of patient-specific strategies and multidisciplinary communication to mitigate the risks of electromagnetic interference from medical equipment in patients with implanted cardiac devices.
Electromagnetic interference produced by medical equipment can interact with implanted cardiac devices such as pacemakers and implantable cardioverter-defibrillators. The most commonly observed interaction is in the operating room with electrosurgery. The risk of interactions can often be mitigated by close communication between the cardiac-device specialist and the anesthesiology/surgical team to develop a patient-specific strategy that accounts for factors such as type of device, type of surgery, and whether the patient is pacemaker dependent. Although magnetic resonance imaging should generally not be used in patients with implanted cardiac devices, several published guidelines provide strategies and recommendations for managing risks if magnetic resonance imaging is required with no suitable diagnostic alternatives. Other common sources of electromagnetic interference in the medical environment are ionizing radiation and left ventricular assist devices.