Answers are grounded in this paper and cite their sources.
Measures refractive index variations in liquid solutions, indicating a reliable sensing platform.
In this paper, we report the design and fabrication of a surface plasmon resonance (SPR) sensor for measuring variations in the refractive index of liquid solutions. The sensor employs a simple Kretschmann configuration in which only two layers of chromium and gold with optimised thicknesses were deposited on a prism and the laser beam reflection is evaluated via image processing instead of photodiode. A computer simulation programme was developed to predict resonance curves by analysing the effect of key parameters. To validate the system, calibration experiments were carried out using sugar solutions with concentrations of 0%, 5%, 15% and 20%. The results show that the implemented SPR sensor achieves a sensitivity of 165.16 deg/RIU, an FWHM of 7 deg, a quality factor of 25.6 RIU −1 and a dynamic range of 1.00–1.37 RIU. These findings demonstrate the reproducibility and robust performance of the proposed setup, highlighting its potential as a low‐cost and reliable platform for refractive index sensing in chemical, biological and industrial applications.
No takes yet. Share an insight, caveat, or question.
Feyzian et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: