Key result
Cardiac motion minimally affects radiation dose distribution, maintaining a gamma pass rate of ~94%.
Why the study?
Cardiac motion may degrade dose distribution during stereotactic arrhythmia radioablation using the CyberKnife robotic radiosurgery system.
Simulated cardiac motion has limited effects on stereotactic arrhythmia radioablation dose distribution, supporting the feasibility of using respiratory motion tracking with an ICD lead as a fiducial marker.
Supports preclinical feasibility of ICD-lead respiratory tracking despite cardiac motion; leaves open human validation for stereotactic arrhythmia radioablation.
PURPOSE: Cardiac motion may degrade dose distribution during stereotactic arrhythmia radioablation using the CyberKnife system, a robotic radiosurgery system. This study evaluated the dose distribution changes using a self-made cardiac dynamic platform that mimics cardiac motion. METHODS: The cardiac dynamic platform was operated with amplitudes of 5 and 3.5 mm along the superior-inferior (SI) and left-right (LR) directions, respectively. The respiratory motion tracking of the CyberKnife system was applied when respiratory motion, simulated using a commercial platform, was introduced. The accuracy of respiratory motion tracking was evaluated by the correlation error between infrared markers and a fiducial marker. The dose distribution was compared with and without cardiac motion. The evaluations included error in the centroid analysis of the irradiated dose distribution, dose profile analysis in the SI and LR directions, and dose distribution analysis comparing the irradiated and planned dose distributions. RESULTS: Cardiac motion increased the correlation error in the direction of motion. Cardiac motion displaced the centroid by up to 0.23 and 0.19 mm in the SI and LR directions, respectively. Cardiac motion blurring caused the distance of the isodose lines to become smaller (bigger) at higher (lower) doses in the SI direction. The gamma pass rate was reduced by cardiac motion but exceeded 94.1% with 1 mm/3% for all conditions. Respiratory motion tracking was also effective under cardiac motion. The cardiac motion slightly varied the dose at the edges of the irradiation volume. CONCLUSION: While cardiac motion increased respiratory tracking correlation errors, its effects on dose distribution were limited in this study. Further studies using motion phantoms that are close to a human or individual patient are necessary for a more detailed understanding of the effects of cardiac motion.
No takes yet. Share an insight, caveat, or question.
Miyachi et al. (2025) studied Ventricular tachycardia (stereotactic arrhythmia radioablation). Cardiac motion simulation vs. No cardiac motion was evaluated on Gamma pass rate for dose distribution. Cardiac motion increased respiratory tracking correlation errors but its effects on dose distribution were limited, with a gamma pass rate exceeding 94.1% with 1 mm/3% criteria for all conditions.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: