Key result
Respiratory beam tracking during STAR is linked to only a ~1% drop in conformity index.
Why the study?
The importance of cardiorespiratory motion during stereotactic arrhythmia radio-ablation (STAR) and its dosimetric impact are still poorly understood.
Does respiratory beam tracking improve dosimetric conformity compared to other motion management strategies during stereotactic arrhythmia radio-ablation in patients with refractory ventricular tachycardia?
Observational (n=61)
Yes
Does respiratory beam tracking improve dosimetric conformity compared to other motion management strategies during stereotactic arrhythmia radio-ablation in patients with refractory ventricular tachycardia?
Respiratory beam tracking significantly limits the dosimetric impact of cardiorespiratory motion during stereotactic arrhythmia radio-ablation compared to other motion management strategies.
Respiratory beam tracking may improve dosimetric conformity in STAR; hypothesis-generating and requires randomized trials before clinical adoption.
Purpose Stereotactic arrhythmia radio-ablation (STAR) is increasingly used to treat refractory ventricular tachycardia. The importance of cardiorespiratory motion during STAR is still poorly understood. We collected multicenter STOPSTORM.eu data to estimate cardiorespiratory motion and its dosimetric impact on STAR-deliveries. Methods and Materials Data were collected from 61 ventricular tachycardia patients. Treatments were performed in free-breathing (FB) with an internal-target volume (FB:ITV) (25 patients), in FB with respiratory beam tracking (FB:TR) (11 patients), using abdominal compression (AC) (10 patients), or in deep inspiration breath-hold (BH) (15 patients). Reference computed tomography images were registered to all breathing/cardiac phases, and the resulting deformation vector fields were used to extract motion for target structures and auto-contoured cardiorespiratory structures (CSs). Additionally, CS were auto-contoured on all breathing/cardiac phases to extract motion independently. Inverse deformation vector fields were used to accumulate motion-adjusted dose distributions. Dose differences were estimated using the conformity index: (V PTV25Gy ) 2 /(V PTV · V 25Gy ), for planning target volume, and maximum and mean dose differences for CS. Results The median 95th percentile planning target volume motion was comparable between patients, with 6.0 mm (FB), 5.6 mm (AC), and 4.9 mm (BH). The left ventricle had the highest median 95th percentile motion out of all CS, with 6.8 mm (FB), 5.5 mm (AC), and 5.9 mm (BH). Estimated motion values were comparable between deformable image registration and auto-contouring (FB, 2.9-7.6 vs 3.1-8.5 mm; AC, 2.4-15.1 vs 3.0-9.0 mm; BH, 1.0-7.3 vs 2.4-6.7 mm). Median conformity index was reduced in the accumulated dose compared to the planned dose by: FB:ITV (–7.2%), AC (–14.1%), BH (–4.3%), and FB:TR (–1.3%). Mean CS dose increased by up to 2 Gy, and maximum dose increased by up to 1.2 Gy (excluding outliers). The largest dose differences were found for small CS. Conclusions Motion and its dosimetric effect have similar magnitudes between treatment setups. Motion findings were consistent between deformable image registration and auto-contouring. Only tracking significantly limited the dosimetric impact of motion.
No takes yet. Share an insight, caveat, or question.
Pol et al. (2025) conducted an observational in refractory ventricular tachycardia (n=61). Stereotactic arrhythmia radio-ablation (STAR) treatment setups was evaluated on Cardiorespiratory motion and dosimetric impact (conformity index). Respiratory beam tracking during stereotactic arrhythmia radio-ablation significantly limited the dosimetric impact of motion, reducing the conformity index by only 1.3% versus 4.3-14.1% in other setups.