Key result
Mid-position planning strategy reduces PTV volume by ~30% compared to internal target volume in STAR.
Why the study?
This study evaluated the benefits of implementing a mid-position approach to compensate for cardio-respiratory motions in stereotactic arrhythmia radioablation compared to an internal target volume approach.
Does the mid-position (MidP) planning target volume strategy reduce planning target volumes and radiation doses to organs-at-risk while maintaining clinical target volume coverage compared to the internal target volume (ITV) approach in patients undergoing stereotactic arrhythmia radioablation for refractory ventricular tachycardia?
Observational (n=15)
No
Does the mid-position (MidP) planning target volume strategy reduce planning target volumes and radiation doses to organs-at-risk while maintaining clinical target volume coverage compared to the internal target volume (ITV) approach in patients undergoing stereotactic arrhythmia radioablation for refractory ventricular tachycardia?
Implementing a mid-position (MidP) motion management strategy for stereotactic arrhythmia radioablation significantly reduces planning target volumes by an average of 30% and lowers radiation exposure to surrounding critical organs (such as the heart, coronary arteries, and stomach) without compromising target dose coverage compared to the standard internal target volume (ITV) approach.
Answers are grounded in this paper and cite their sources.
Supports mid-position PTV for smaller volumes and lower cardiac doses in arrhythmia radioablation; hypothesis-generating, should not yet change practice.
Purpose/objective This study proposes an implementation of the mid‐position (MidP) approach to compensate for cardio‐respiratory motions in the context of Stereotactic Arrhythmia Radioablation (STAR) and evaluates its benefits compared to an internal target volume (ITV) approach. Materials and methods Fifteen patients who underwent STAR for refractory ventricular tachycardia in our institution were included in this retrospective planning study. For each patient, a cardiac‐gated four‐dimensional computed tomography (4D‐CT card ) scan and a respiratory‐gated four‐dimensional computed tomography (4D‐CT resp) were acquired. All patients were treated using a volumetric modulated arc therapy technique using an in‐treatment Cone‐Beam CT (CBCT) image guidance. The MidP approach was implemented to compensate for uncertainties, including cardio‐respiratory motions characterized using the 4D‐CT card and 4D‐CT resp scans, and the inter‐fraction motions measured using the CBCT scans. For comparison purposes, the ITV approach was also implemented. Both approaches were compared in terms of planning target volume (PTV) volumes, doses to organs‐at‐risk, and clinical target volume (CTV) doses, assessed using a 4D modeling method that estimates the accumulated dose. Results Compared with the ITV method, the MidP approach resulted in a mean [min‐max] relative PTV volume reduction of 30% [19%, 48%] ( p < 0.001, Wilcoxon signed rank test). The mean [min‐max] D95% CTV coverage was 105% [101%–114%] and 107% [101%–117%] of the prescription dose for MidP and ITV‐based plans, respectively. The median dose to the whole heart was significantly lower with MidP‐based plans with a mean difference of −0.5 Gy ( p = 0.0084). The near‐maximum dose (D1%) delivered to left coronary arteries, aorta, and stomach was systematically lower with the MidP‐based plans. Conclusion Compared to ITV based approach, the use of MidP strategy for treatment planning of STAR leads to significantly smaller PTV and lower surrounding OAR doses while still achieving a clinically acceptable CTV coverage.
No takes yet. Share an insight, caveat, or question.
Gabillaud et al. (2025) conducted an observational in Refractory ventricular tachycardia (n=15). Mid-position (MidP) planning target volume (PTV) strategy vs. Internal target volume (ITV) PTV strategy was evaluated on Relative planning target volume (PTV) reduction (Mean reduction 30%, p=<0.001). Mid-position planning target volume strategy reduced PTV volume by a mean of 30% and lowered whole-heart median dose by 0.5 Gy compared to the internal target volume approach.
Synapse has enriched 10 related papers. Consider them for comparative context: