Answers are grounded in this paper and cite their sources.
Bacillus subtilis var. niger spores were used to determine the exposure time for formaldehyde decontamination of biological safety cabinets. Formaldehyde contact times less than 3 hr were insufficient for sterilization. A contact time of 4 hr or more resulted in a reproducible killing of the spore strips placed inside the cabinets. At 6 hr sufficient formaldehyde had diffused through the high efficiency particulate air (HEPA) filter to sterilize the strips with lower spore counts. A minimum of 5 to 6 logs of killing were observed after 4 to 6 hr of treatment.
No takes yet. Share an insight, caveat, or question.
Fink et al. (1988) studied this question.