Key result
Surpoint Tag Index showed a weak but positive correlation with cavotricuspid isthmus thickness in ablation lesions exhibiting bidirectional block.
Why the study?
The relationship between the Surpoint Tag Index and in-vivo atrial wall thickness in patients exhibiting bidirectional cavotricuspid isthmus block remained unknown.
Observational (n=25)
Single-blind
A mean Tag Index value of 455 correlates with bidirectional block across the bulk of the cavotricuspid isthmus, though tissue composition also influences the energy required.
Answers are grounded in this paper and cite their sources.
Tag Index-CTI thickness association is hypothesis-generating for ablation lesion assessment; prospective validation needed before clinical use.
Background An RFA lesion quality indicator, Surpoint Tag Index ® (TI) incorporates key factors: power, time, and contact force, impacting lesion quality. TI accurately estimates lesion depth in animal studies. However, the relationship between TI and in‐vivo atrial wall thickness in patients exhibiting bidirectional block remains unknown. Objective To describe the relationship between atrial wall thickness and TI in CTI exhibiting bidirectional block. Methods Data from 492 RFA lesions from 25 patients undergoing PVI and CTI ablations in SR with point‐by‐point RF lesions (<45 W) utilizing a Thermocool Smarttouch ® SF ablation catheter and CARTO‐3 mapping were retrospectively analyzed. Operators were blinded to TI data and CTI thickness. CTI thickness was obtained using ICE images on Cartosound pre‐ablation. Durable lesions were defined as part of a lesion set exhibiting bidirectional block of >30 min. Results In lesions exhibiting bidirectional block, the thinnest (1–2 mm; 5% lesions) and thickest (8–10 mm; 6% lesions) portions of the CTI correlated with the lowest (429 ± 75) and highest (516 ± 64) TI. The bulk of thickness (2–6 mm; 80%) correlated with a TI of 455 ± 72 ( p = 0.001). There was a weak but positive correlation between TI and CTI thickness ( r = 0.2; p ≤ 0.01). Examined in sectors, the anterior 1/3 rd CTI was the thickest (4.8 ± 1.9 mm) but correlated with a similar TI value (479 ± 75 vs. 471 ± 70; p = 0.34) as the thinner middle 1/3 rd (3.8 ± 1.7 mm; p ≤ 0.0001). Conclusion A mean TI value of 455 correlates with bidirectional block across the bulk of CTI with lower and higher values needed for the thinner and thicker portions, respectively. Tissue composition, aside from wall thickness, influences TI values for the creation of the bidirectional block.
No takes yet. Share an insight, caveat, or question.
Smith et al. (2021) conducted an observational in Patients undergoing PVI and CTI ablations (n=25). Surpoint Tag Index (TI) was evaluated on Correlation between TI and CTI thickness in lesions exhibiting bidirectional block (r = 0.2, p=≤0.01). Surpoint Tag Index showed a weak but positive correlation with cavotricuspid isthmus thickness in ablation lesions exhibiting bidirectional block.
Synapse has enriched 10 related papers. Consider them for comparative context: