Key result
Closed cooled-tip and 8-mm-tip catheters demonstrated equal and high efficacy for radiofrequency ablation of common atrial flutter, achieving complete bidirectional isthmus block in 99% of patients.
Why the study?
Does radiofrequency ablation with a closed cooled-tip catheter improve bidirectional isthmus conduction block compared to an 8-mm-tip catheter in patients with common atrial flutter?
RCT (n=100)
randomized
Does radiofrequency ablation with a closed cooled-tip catheter improve bidirectional isthmus conduction block compared to an 8-mm-tip catheter in patients with common atrial flutter?
Closed cooled-tip and 8-mm-tip catheters demonstrate equal and high efficacy for radiofrequency ablation of common atrial flutter.
Answers are grounded in this paper and cite their sources.
Both catheter types offer reliable options now; confirms equivalence for typical flutter ablation.
INTRODUCTION: Cooled-tip and 8-mm-tip catheters have been found to be more effective than conventional 4-mm-tip catheters for radiofrequency (RF) ablation of common atrial flutter. The aim of this study was to compare the efficacy and safety of cooled-tip and 8-mm-tip catheters for flutter ablation in a randomized, prospective study. METHODS AND RESULTS: In 100 consecutive patients referred for ablation of common atrial flutter, cavotricuspid ablation was performed with a closed cooled-tip catheter (n = 50) or an 8-mm-tip ablation catheter (n = 50). RF current was applied for 60 to 120 seconds at powers of 40 to 50 W with the closed cooled-tip catheter and in a temperature-controlled mode (65 degrees C/70 W) with the 8-mm-tip catheter. The endpoint was achievement of a bidirectional isthmus conduction block. Cross-over was performed after 15 unsuccessful RF applications for each of the catheters. Complete bidirectional isthmus block was achieved in 99% of patients. Cross-over was performed in 11 patients after primary use of the cooled-tip catheter and in 9 patients after primary ablation with the 8-mm-tip catheter. No significant differences were found in the procedure parameters, such as overall RF applications (12.4 +/- 11.3 vs 12.9 +/- 8.6), ablation duration (42 +/- 43 min vs 39 +/- 27 min), and fluoroscopy time (17.0 +/- 18.7 min vs 15.7 +/- 10.7 min). In a mean follow-up of 8.3 months, 1 patient in the cooled-tip group and 3 patients in the 8-mm-tip group had recurrence of common atrial flutter. CONCLUSION: Use of the closed cooled-tip ablation catheter and the 8-mm-tip catheter have equal and high efficacy for RF ablation of common atrial flutter.
No takes yet. Share an insight, caveat, or question.
Schreieck et al. (2002) conducted an RCT in Common atrial flutter (n=100). Closed cooled-tip catheter vs. 8-mm-tip ablation catheter was evaluated on Achievement of a bidirectional isthmus conduction block. Closed cooled-tip and 8-mm-tip catheters demonstrated equal and high efficacy for radiofrequency ablation of common atrial flutter, achieving complete bidirectional isthmus block in 99% of patients.