Answers are grounded in this paper and cite their sources.
Steam‐sterilizable membrane probes for monitoring the dissolved oxygen level in fermentors, or the oxygen content of gas streams, are described. The probes have a silver cathode, a lead anode, and an acetate buffer as an electrolyte. The membrane is Teflon. The current output of the probes in the absence of oxygen is negligible.
No takes yet. Share an insight, caveat, or question.
Johnson et al. (1964) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: