Key result
Novel NRMS classifier detects ventricular ectopic beat episodes with ~100% accuracy.
Why the study?
Manual cross-checking and annotation of long-term Holter ECG records takes considerable time, requiring an effective automatic cardiac episode detection technique to reduce manual burden.
Does an automated signal processing framework using an NRMS classifier accurately detect ventricular ectopic beats in long-term ECG monitoring?
Does an automated signal processing framework using an NRMS classifier accurately detect ventricular ectopic beats in long-term ECG monitoring?
An automated cross-database NRMS machine learning model can accurately and rapidly classify normal and ventricular ectopic beats in long-term ECGs.
May aid automated VEB detection in long-term ECGs; leaves open prospective clinical validation before practice adoption.
The Holter monitor captures the electrocardiogram (ECG) and detects abnormal episodes, but physicians still use manual cross-checking. It takes a considerable time to annotate a long-term ECG record. As a result, research continues to be conducted to produce an effective automatic cardiac episode detection technique that will reduce the manual burden. The current study presents a signal processing framework to detect ventricular ectopic beat (VEB) episodes in long-term ECG signals of cross-database. The proposed study has experimented with the cross-database of open-source and proprietary databases. The ECG signals were preprocessed and extracted the features such as pre-RR interval, post-RR interval, QRS complex duration, QR slope, and RS slope from each beat. In the proposed work, four models such as support vector machine, k-means nearest neighbor, nearest mean classifier, and nearest RMS (NRMS) classifiers were used to classify the data into normal and VEB episodes. Further, the trained models were used to predict the VEB episodes from the proprietary database. NRMS has reported better performance among four classification models. NRMS has shown the classification accuracy of 98.68% and F1-score of 94.12%, recall rate of 100%, specificity of 98.53%, and precision of 88.89% with an open-source database. In addition, it showed an accuracy of 99.97%, F1-score of 94.54%, recall rate of 98.62%, specificity of 99.98%, and precision of 90.79% to detect the VEB cardiac episodes from the proprietary database. Therefore, it is concluded that the proposed framework can be used in the automatic diagnosis system to detect VEB cardiac episodes.
No takes yet. Share an insight, caveat, or question.
Srinivasulu et al. (2023) studied Patients undergoing long-term ECG monitoring (n=30). Nearest RMS (NRMS) machine learning classifier vs. Support vector machine, KNN, nearest mean classifier, and manual cardiologist annotation was evaluated on Detection accuracy of ventricular ectopic beat (VEB) cardiac episodes from a proprietary cross-database. The nearest RMS (NRMS) classifier achieved a classification accuracy of 99.97% and an F1-score of 94.54% for detecting ventricular ectopic beat episodes from a proprietary cross-database.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: