BACKGROUND: Balancing reduction accuracy with soft-tissue preservation is a challenge in orthopaedics. Computer-assisted orthopaedic surgery (CAOS) can improve accuracy and reduce radiation exposure. However, previous reports have not summarized the fracture patterns to which CAOS has been applied. METHODS: We used a CAOS system and a stereolithography model to define a new fracture classification. Twenty reduction tests were performed to evaluate the effectiveness of preoperative trajectory planning. RESULTS: Twenty tests ran automatically and smoothly. Only three slight scratches occurred. Seventy-six path points represented displacement deviations of < 2 mm (average < 1 mm) and angulation deviation of < 1.5°. DISCUSSION: Because of the strength of muscles, mechanical sensors are used to prevent iatrogenic soft-tissue injury. Secondary fractures are prevented mainly through preoperative trajectory planning. Based on our data, a 1 mm gap between the edges of fractures spikes is sufficient to avoid emergency braking from spike interference.
No takes yet. Share an insight, caveat, or question.
Du et al. (2014) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: