Key result
Real-time MRI-guided right heart catheterization matches X-ray guidance with ~94% procedural success.
Why the study?
Does real-time MRI-guided right heart catheterization maintain procedural success and time compared to X-ray guidance in adults undergoing clinical cardiac catheterization?
Does real-time MRI-guided right heart catheterization maintain procedural success and time compared to X-ray guidance in adults undergoing clinical cardiac catheterization?
Absolute Event Rate: 93.75% vs 93.75%
Real-time MRI-guided right heart catheterization using gadolinium-filled balloon catheters is feasible and comparable to X-ray guidance in procedural success and time.
AIMS: Real-time MRI creates images with superb tissue contrast that may enable radiation-free catheterization. Simple procedures are the first step towards novel interventional procedures. We aim to perform comprehensive transfemoral diagnostic right heart catheterization in an unselected cohort of patients entirely using MRI guidance. METHODS AND RESULTS: We performed X-ray and MRI-guided transfemoral right heart catheterization in consecutive patients undergoing clinical cardiac catheterization. We sampled both cavae and both pulmonary arteries. We compared success rate, time to perform key steps, and catheter visibility among X-ray and MRI procedures using air-filled or gadolinium-filled balloon-tipped catheters. Sixteen subjects (four with shunt, nine with coronary artery disease, three with other) underwent paired X-ray and MRI catheterization. Complete guidewire-free catheterization was possible in 15 of 16 under both. MRI using gadolinium-filled balloons was at least as successful as X-ray in all procedure steps, more successful than MRI using air-filled balloons, and better than both in entering the left pulmonary artery. Total catheterization time and individual procedure steps required approximately the same amount of time irrespective of image guidance modality. Catheter conspicuity was best under X-ray and next-best using gadolinium-filled MRI balloons. CONCLUSION: In this early experience, comprehensive transfemoral right heart catheterization appears feasible using only MRI for imaging guidance. Gadolinium-filled balloon catheters were more conspicuous than air-filled ones. Further workflow and device enhancement are necessary for clinical adoption.
No takes yet. Share an insight, caveat, or question.
Ratnayaka et al. (2012) studied Patients undergoing clinical cardiac catheterization (n=16). MRI-guided transfemoral right heart catheterization vs. X-ray-guided transfemoral right heart catheterization was evaluated on Complete guidewire-free catheterization. Real-time MRI-guided transfemoral right heart catheterization was feasible, achieving complete guidewire-free catheterization in 15 of 16 patients (93.8%), identical to X-ray guidance.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: