Key result
A high-input-impedance bioimpedance sensor interface IC achieved an input impedance of 100 MΩ at 50 kHz, 0.5-mΩ/√Hz sensitivity at 1 Hz, and a 106-dB SNR for cardio-respiratory monitoring.
Why the study?
Small-area dry-electrode cardio-respiratory signal monitoring requires high-precision bioimpedance sensing capable of overcoming high-impedance dry electrodes.
A novel bioimpedance sensor interface IC enables high-precision gel-free cardio-respiratory monitoring with high input impedance and signal-to-noise ratio.
This article describes a high-input-impedance, low-noise bioimpedance (BioZ) sensor interface IC for small-area dry-electrode cardio-respiratory signals monitoring. To facilitate high-precision BioZ sensing with high-impedance dry electrodes, the IC utilizes three key techniques as follows: 1) a bias control loop (BCL) to eliminate the excitation current mismatch, reducing the voltage fluctuation on high-impedance input; 2) a quiet-chopping current feedback instrumentation amplifier (QC-CFIA) to mitigate the input-signal-dependent noise; and 3) a full pre-charge (FPC) technique to cancel the input parasitic capacitance for impedance boosting. Manufactured in a 0.18- <tex-math notation="LaTeX">μ m </tex-math> CMOS process, the BioZ prototype IC occupies an area of 0.4 mm2 while consuming 15.8- and 14.4–128.4- <tex-math notation="LaTeX">μ W </tex-math> current, respectively, from the amplifier and the excitation current generator (CG). With these proposed techniques, this IC achieves a high input impedance of 100 <tex-math notation="LaTeX">MΩ </tex-math> at 50 kHz, 0.5- <tex-math notation="LaTeX">mΩ / </tex-math> Hz sensitivity at 1 Hz, and a 106-dB signal-to-noise ratio (SNR). A gel-free respiration and impedance cardiography (ICG) recording has been successfully demonstrated on the human body with four 0.45-cm2 dry electrodes.
No takes yet. Share an insight, caveat, or question.
Pan et al. (2022) studied this question. BioZ sensor interface IC was evaluated on Input impedance, sensitivity, and signal-to-noise ratio. A high-input-impedance bioimpedance sensor interface IC achieved an input impedance of 100 MΩ at 50 kHz, 0.5-mΩ/√Hz sensitivity at 1 Hz, and a 106-dB SNR for cardio-respiratory monitoring.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: