Does a sticker-type ECG/PPG concurrent monitoring system accurately measure pulse arrival time and SpO2 compared to a reference device?
Does a sticker-type ECG/PPG concurrent monitoring system accurately measure pulse arrival time and SpO2 compared to a reference device?
A novel, lightweight, low-power sticker-type ECG/PPG sensor system demonstrated feasibility for concurrent monitoring of pulse arrival time and SpO2 in a small human cohort.
Answers are grounded in this paper and cite their sources.
Supports feasibility of unobtrusive wearable monitoring; leaves open larger validation trials before clinical adoption.
The sticker-type sensor system is proposed targeting ECG/PPG concurrent monitoring for cardiovascular diseases. The stickers are composed of two types: Hub and Sensor-node (SN) sticker. Low-power CMOS SoC for measuring ECG and PPG signal is hybrid integrated with organic light emitting diodes (OLEDs) and organic photo detector (OPD). The sticker has only 2g weight and only consumes 141μW. The optical calibration loop is adopted for maintaining SNR of PPG signal higher than 30dB. The pulse arrival time (PAT) and SpO2 value can be extracted from various body parts and verified comparing with the reference device from 20 people in-vivo experiments.