Key result
A portable 5-GHz radar system for noncontact heartbeat and respiration monitoring achieved a detection accuracy of >80% at a distance of 2.8 m and transmitted power of 20 muW.
A compact, low-cost 5-GHz radar system can accurately monitor heartbeat and respiration without contact at a distance of 2.8 m.
Supports portable noncontact vital-signs radar feasibility; leaves open clinical validation before practice adoption.
A portable noncontact heartbeat and respiration monitoring system operating in 5-GHz band is reported in this letter. Compared with the previously reported system operating at Ka-band, this system has been simplified to include only two PCB antennas, a palm-size PCB radio module, a data acquisition module, and a laptop. The system, powered by the laptop battery through USB connection, is compact, low-cost, and convenient for field test. A detection accuracy of better than 80% has been achieved at a distance of 2.8 m and at a low transmitted power of 20 muW. The system can be used for various applications in biology, medicine, and security.
No takes yet. Share an insight, caveat, or question.
Xiao et al. (2007) studied Heartbeat and respiration monitoring. Portable noncontact heartbeat and respiration monitoring system using 5-GHz radar vs. Previously reported Ka-band system was evaluated on Detection accuracy at a distance of 2.8 m and transmitted power of 20 muW. A portable 5-GHz radar system for noncontact heartbeat and respiration monitoring achieved a detection accuracy of >80% at a distance of 2.8 m and transmitted power of 20 muW.