Key result
Five blood glucose monitoring systems using Verio test strips met or exceeded ISO15197:2013(E) accuracy standards for user performance and system accuracy in a clinical setting.
Why the study?
Do blood glucose monitoring systems using Verio test strips meet ISO15197:2013(E) accuracy and precision standards?
Do blood glucose monitoring systems using Verio test strips meet ISO15197:2013(E) accuracy and precision standards?
Answers are grounded in this paper and cite their sources.
The evaluated blood glucose monitoring systems using Verio test strips demonstrate acceptable accuracy and precision according to ISO standards.
Accurate self-monitoring of blood glucose is a key component of effective self-management of glycemic control. Accurate self-monitoring of blood glucose results are required for optimal insulin dosing and detection of hypoglycemia. However, blood glucose monitoring systems may be susceptible to error from test strip, user, environmental and pharmacological factors. This report evaluated 5 blood glucose monitoring systems that each use Verio glucose test strips for precision, effect of hematocrit and interferences in laboratory testing, and lay user and system accuracy in clinical testing according to the guidelines in ISO15197:2013(E). Performance of OneTouch(®) VerioVue™ met or exceeded standards described in ISO15197:2013 for precision, hematocrit performance and interference testing in a laboratory setting. Performance of OneTouch(®) Verio IQ™, OneTouch(®) Verio Pro™, OneTouch(®) Verio™, OneTouch(®) VerioVue™ and Omni Pod each met or exceeded accuracy standards for user performance and system accuracy in a clinical setting set forth in ISO15197:2013(E).
No takes yet. Share an insight, caveat, or question.
Katz et al. (2015) studied Blood glucose monitoring. Blood glucose monitoring systems using Verio test strips vs. ISO15197:2013(E) standards was evaluated on Precision, effect of hematocrit, interferences, lay user and system accuracy according to ISO15197:2013(E). Five blood glucose monitoring systems using Verio test strips met or exceeded ISO15197:2013(E) accuracy standards for user performance and system accuracy in a clinical setting.