Answers are grounded in this paper and cite their sources.
The Oxylog is a portable instrument designed to measure the oxygen consumption ([Vdot]02) an ambulatory subject. Steady-state measurements have been made, using an Oxylog, of inspiratory volume ( [Vdot]1) [Vdot]02 during bicycle ergometer exercise at work rates ranging from 30 to 150 W. These measurements have been compared with simultaneous measurements of expiratory volume ([Vdot] E) and [Vdot]O2 made using a dry gas meter and mass spectrometer. Four experiments were conducted, during which a total of 433 comparative measurements were made. In two experiments the Oxylog significantly underestimated [Vdot]O2 (by 4-4 and 5-6%). Averaging over the four experiments, however, the underestimate reduced to 1-5%, which could be accounted for by a respiratory exchange ratio of approximately 0-9. There was, overall, no significant difference between [Vdot] 1 and [Vdot] E. It is concluded that the Oxylog is sufficiently accurate for the reliable determination of [Vdot]02 and of energy expenditure under field conditions.
No takes yet. Share an insight, caveat, or question.
Harrison et al. (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: