Key result
Intra-operative balloon pressure-volume sizing accurately determines annular diameter with high repeatability in preclinical models.
Why the study?
Preoperative image-based sizing for TAVI has inherent limitations leading to suboptimal prosthesis sizing and complications such as aortic regurgitation or conduction disturbance.
Does an intra-operative sizing method using balloon pressure-volume data accurately determine the dimension of compliant aortic annuli in idealized phantoms?
Does an intra-operative sizing method using balloon pressure-volume data accurately determine the dimension of compliant aortic annuli in idealized phantoms?
An intra-operative sizing algorithm using balloon pressure-volume data can accurately determine the dimensions of compliant and elliptical aortic annuli in vitro, representing progress toward improving TAVI device selection.
May aid TAVI sizing if validated in vivo; leaves open translation from phantoms to patients.
OBJECTIVE: Aortic stenos (AS) is a heart valve disease that commonly affects the elderly. Transcatheter aortic valve implantation is a minimally invasive treatment that allows to replace the function of the diseased native valve with a prosthetic device, relying on catheters for device implantation. According to the current clinical guidelines, the choice of the implanted device is based on preoperative sizing determined by image-based technology. However, this assessment faces inherent limitations that can lead to sub-optimal sizing of the prosthesis; in turn, this can cause major post-operative complications like aortic regurgitation or cardiac electrical signal disruption. METHOD: By utilizing balloon pressure and volume data, this article proposes an intra-operative method for determining the dimension of the aortic annulus which takes into account its compliance and geometric irregularity. The intra-balloon pressure-volume curves were obtained using an Automated Balloon Inflation Device operating a commercially available valvuloplasty balloon catheter. A sizing algorithm to estimate the dimensions of the annulus was integrated via a validated analytical model and a numerical model for balloon free-inflation. Tests were performed on circular and elliptical idealised aortic phantoms. RESULTS: Experimental results confirm that the pressure-volume data processed with the sizing algorithm can be used to determine the circular annular diameter for all tissue rigidities. CONCLUSION: The measurement of stiffer elliptical annulus phantoms shows good accuracy and high repeatability. SIGNIFICANCE: This work represents substantial progress toward improving the selection of TAVI devices by using balloon catheters to improve the sizing of compliant aortic annuli with complex geometry.
No takes yet. Share an insight, caveat, or question.
Yao et al. (2023) studied Aortic Stenosis (in vitro model) (n=32). Automated Balloon Inflation Device (ABID) sizing algorithm vs. Known phantom dimensions was evaluated on Estimated annular diameter. The intra-operative sizing algorithm using balloon pressure-volume data accurately determined the circular annular diameter across all tissue rigidities and demonstrated high repeatability in stiffer elliptical phantoms.