Develops a Cd detection kit for water safety, suggesting improved monitoring solutions.
Access to clean drinking water is a fundamental human need that can be sourced from rivers such as the Citarum River in West Java, Indonesia. However, according to a 2018 World Bank report, the Citarum River is heavily contaminated with cadmium (Cd), with concentrations ranging from 0.03 mg/L to 0.464 mg/L—far exceeding the Indonesian water quality standard of 0.01 mg/L (10 ppb) stipulated in Government Regulation No. 22/2021. To enable the detection of Cd at low concentrations, a portable detection kit based on the Cyclic Voltammetry (CV) method was developed. The system integrates an LMP91000 potentiostat, Arduino UNO microcontroller, ADS1115 analog-to-digital converter, and a Nextion NX8048K050 HMI display. A Glassy Carbon Electrode (GCE) with a graphene-doped ZnO/PVA nanocomposite as the working electrode was employed to enhance sensitivity. The developed kit demonstrated a sensitivity of 4.92 μA/ppb, a linearity (R 2 ) of 0.96, and a limit of detection (LOD) of 4.08 ppb. Measurement errors ranged from 2.26% to 5.02%, confirming that the proposed system can effectively detect Cd concentrations around the regulatory threshold of 10 ppb.
No takes yet. Share an insight, caveat, or question.
Ismardi et al. (2026) studied this question.