Key result
ST Genesia thrombin generation testing shows high reproducibility with <10% variation in healthy adults.
Why the study?
Fully automated thrombin generation assays facilitate transition to clinical use, but assay performance and reference intervals needed evaluation across laboratories using the ST Genesia analyzer.
Observational (n=252)
Yes
The ST Genesia platform provides highly reproducible automated thrombin generation measurements across multiple laboratories, supporting its standardized use in clinical and research settings.
High interlaboratory reproducibility supports thrombin generation assay standardization; leaves open impact on clinical decision-making.
Introduction Fully automated thrombin generation (TG) assays facilitate the transition from research to clinical application. We conducted a multicenter inter‐laboratory study using the ST Genesia analyzer to evaluate assay performance and establish reference intervals in healthy adults from Spain and Portugal. Methods In this multicenter study, plasma samples from healthy adult volunteers were analyzed across 12 laboratories following a harmonized protocol. TG was measured using the STG‐ThromboScreen reagent in the presence and absence of thrombomodulin (TM). Analytical performance and reference intervals were assessed. Results Within‐run and intra‐day precision were excellent (coefficient of variation < 10%) for all TG parameters except velocity index. Normalization reduced interlaboratory variability for amplitude‐based parameters such as endogenous thrombin potential (ETP) and peak height, whereas its effect was less pronounced for velocity index. Satisfactory interlaboratory performance (| z | ≤ 2) was achieved in 96%–98% of results, and normalized parameters showed minimal bias across centers. In 252 healthy adults, median normalized values were: ETP 131% (IQR 118–142), peak height 116% (IQR 104–127), lag time 2.6 min (IQR 2.3–2.8), and time‐to‐peak 5.5 min (IQR 5.0–5.9). Median ETP inhibition in the presence of TM was 47.2% (IQR 35.1–61.8). Age had a modest influence on TG parameters, whereas sex differences were observed. Conclusions This multicenter interlaboratory study in the Iberian Peninsula demonstrates high reproducibility of TG measurement using the ST Genesia platform with the STG‐ThromboScreen reagent. These findings support its use for standardized TG assessment and provide reference intervals for clinical and research settings.
No takes yet. Share an insight, caveat, or question.
Velasco‐Rodríguez et al. (2026) conducted an observational in Healthy adults (n=252). ST Genesia analyzer with STG-ThromboScreen reagent was evaluated on Analytical performance and reference intervals. Thrombin generation measurement using the ST Genesia platform with the STG-ThromboScreen reagent demonstrated high reproducibility (CV < 10%) in 252 healthy adults.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: