Randomized trial demonstrates new biosensor monitors barrier integrity in cell cultures, suggesting broader research accessibility.
Real-time monitoring of biological barrier integrity is crucial for drug development and disease modeling. The gold-standard technique, transendothelial electrical resistance (TEER), is often limited by the cost and complexity of commercial alternating current systems. To address this, we developed a novel, low-cost biosensor based on a direct current (DC) series voltage division principle, featuring custom hardware and open-source firmware. Validation demonstrated a wide dynamic range (155-105 600 Ω cm2) and high accuracy (±3%). The device showed excellent correlation with a commercial EVOM3 system in monitoring TEER trends during endothelial barrier formation and oxidative stress-induced disruption. Biosensor readings were consistent with barrier kinetics captured by xCelligence RTCA and live-cell imaging. Furthermore, a strong negative correlation was established between decreasing TEER values and increasing paracellular leakage of sodium fluorescein. These results collectively validate our DC-based system as a reliable, accurate, and accessible tool for quantifying in vitro barrier integrity, with significant potential to democratize research in biomedicine and toxicology.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: