Methodology outlines the testing process for solid oxide electrolysis cells, ensuring consistent performance evaluation.
Presented is a simple, step-by-step methodology describing how to assemble, seal, precondition, and test planar 5 × 5 or 10 × 10 cm 2 electrode-supported solid oxide cells (SOCs) for operating in electrolysis mode as solid oxide electrolysis cells (SOECs). Careful control of the gas environment, temperature, quality of sealing, and consistency in attaching the electrical contacts are essential for establishing a representative baseline electrochemical performance of SOECs and obtaining reproducible results when testing multiple cells and when comparing the performance of different SOEC components. This becomes more important when comparing the performances of SOEC produced and tested by the various research groups. Recommendations on the final steps before test termination and disassembling to preserve cell chemical and structural integrity for post-mortem analyses are included.