Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
June 14, 2026Medical PhysicsOpen Access

Development and validation of a novel pulse optimization and beam control system for conventional and ultra high dose‐per‐pulse (FLASH) irradiation

View Full Paper
Ask AI
Bookmark
Share

Authors

LCLuke ConnellNENolan EsplenRLRebecca Lim

Discussion

Loading...

Member takes

Overview

Randomized trial demonstrates enhanced dose precision in FLASH radiotherapy, implying improved treatment outcomes.

Key Points

  • To develop and validate an external beam control system (EBCS) for precise dose control in FLASH radiotherapy.
  • Developed an EBCS interfaced with clinical linear accelerators for conventional and UH-DPPs.
  • Utilized a beam current transformer and an internal transmission ion chamber for dose monitoring.
  • Tested performance by measuring beam-off latency and output stability under various conditions.
  • Achieved a beam-off latency of 56.7 µs (± 4.9 µs).
  • Reduced dose per pulse variability from 26.7% to less than 0.5% with radiofrequency optimization.
  • Total output decreased by up to 20% when voltage inputs deviated from optimal values beyond ± 10%.

Cite This Study

Connell et al. (2026) studied this question.

synapsesocial.com/papers/6a2e46dbb1cc60ccdea8b826https://doi.org/10.1002/mp.70522
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Development and evaluation of an in vivo dose‐based monitoring system for electron FLASH radiation therapy2026
  2. 2Beam intensity and stability control on a modified clinical linear accelerator for FLASH irradiation2025
  3. 3Electron FLASH platform for pre-clinical research: LINAC modification, simplification of pulse control and dosimetry2024 · 1 citations
  4. 4Ultra-high dose rate electron FLASH beam irradiation using a modified clinical linear accelerator2026