Why the study?
Heart rate variability determines autonomic nervous system activity and cardiovascular risk, but the validity and reliability of commercial finger photoplethysmography systems in pediatric real-world settings needed evaluation.
Does a commercially available finger photoplethysmography (PPG) system validly and reliably measure heart rate variability compared to the Polar H10 sensor in healthy children?
Does a commercially available finger photoplethysmography (PPG) system validly and reliably measure heart rate variability compared to the Polar H10 sensor in healthy children?
The Polar H10 heart rate sensor, but not a commercially available finger PPG system, provides reliable heart rate variability measurements in young children in real-world educational settings.
Answers are grounded in this paper and cite their sources.
Finger PPG lacks validity for HRV assessment in young children; small cross-sectional data leaves Polar H10 reliability open to prospective confirmation.
Heart rate variability (HRV) is an accepted method for determining autonomic nervous system activity and cardiovascular risk in various populations. This study assessed the validity and reliability of a commercially available finger photoplethysmography (PPG) system for measuring pediatric HRV in a real-world setting. Sixteen healthy children (4.06 ± 0.58 years) were recruited. The PPG system was compared to the Polar H10 heart rate (HR) sensor validated against ECG (gold standard) for HRV measurement. Seated short-term resting R-R intervals were recorded simultaneously using both systems. Recordings were performed on 3 days at the participants' school. Paired t-tests, effect sizes and Bland-Altman analyses determined the validity of the PPG system. The relative and absolute reliability of both systems were calculated. No HRV parameters were valid for the PPG system. Polar H10 yielded moderate (0.50-0.75) to good (0.75-0.90) relative reliability with R-R intervals and the standard deviation of instantaneous and continuous R-R variability ratio showing the best results (ICCs = 0.84). Polar H10 displayed better absolute reliability with the root mean square of successive differences, R-R intervals and HR showing the lowest values (TEM% < 12%). The use of the Polar H10 and not the PPG system is encouraged for HRV measurement of young children in an educational real-world setting.
No takes yet. Share an insight, caveat, or question.
Speer et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: