Key result
Computer-assisted auscultation reliably captures carotid vascular sounds concentrated between 3 and 40 Hz.
Why the study?
Carotid auscultation is a simple method to screen for atherosclerotic disease and bruits, but it is currently subjective and examiner dependent, while standard imaging modalities require costly hardware.
The proposed device provides a stable, easy-to-use platform for reproducible recording of carotid vascular sounds, paving the way for low-cost, sound-based vascular monitoring.
Supports device refinement for carotid sound recording; leaves open diagnostic utility in vascular disease.
INTRODUCTION: Atherosclerotic diseases of the carotid are a primary cause of cerebrovascular events such as stroke. For the diagnosis and monitoring angiography, ultrasound- or magnetic resonance-based imaging is used which requires costly hardware. In contrast, the auscultation of carotid sounds and screening for bruits - audible patterns related to turbulent blood flow - is a simple examination with comparably little technical demands. It can indicate atherosclerotic diseases and justify further diagnostics but is currently subjective and examiner dependent. METHODS: We propose an easy-to-use computer-assisted auscultation system for a stable and reproducible acquisition of vascular sounds of the carotid. A dedicated skin-transducer-interface was incorporated into a handheld device. The interface comprises two bell-shaped structures, one with additional acoustic membrane, to ensure defined skin contact and a stable propagation path of the sound. The device is connected wirelessly to a desktop application allowing real-time visualization, assessment of signal quality and input of supplementary information along with storage of recordings in a database. An experimental study with 5 healthy subjects was conducted to evaluate usability and stability of the device. Five recordings per carotid served as data basis for a wavelet-based analysis of the stability of spectral characteristics of the recordings. RESULTS: The energy distribution of the wavelet-based stationary spectra proved stable for measurements of a particular carotid with the majority of the energy located between 3 and 40 Hz. Different spectral properties of the carotids of one individual indicate the presence of sound characteristics linked to the particular vessel. User-dependent parameters such as variations of the applied contact pressure appeared to have minor influence on the general stability. CONCLUSION: The system provides a platform for reproducible carotid auscultation and the creation of a database of pathological vascular sounds, which is a prerequisite to investigate sound-based vascular monitoring.
No takes yet. Share an insight, caveat, or question.
Sühn et al. (2020) studied Healthy volunteers (n=5). Computer-assisted auscultation system (CAAS) was evaluated on Spectral stability and reproducibility of acquired vascular sounds. The computer-assisted auscultation system demonstrated stable and reproducible acquisition of carotid vascular sounds, with the majority of the spectral energy located between 3 and 40 Hz.