Key result
Compensatory reserve measurement demonstrated superior accuracy (ROC AUC 0.93) compared with heart rate variability (AUC ≤0.84) and complexity (AUC ≤0.86) for detecting hemodynamic decompensation.
Why the study?
The added benefit of heart rate variability or heart rate complexity for decision support in trauma monitoring remains unclear compared to compensatory reserve measurement.
Does compensatory reserve measurement (CRM) provide greater sensitivity and specificity than heart rate variability (HRV) and heart rate complexity (HRC) to detect progressive reductions in central circulating blood volume and onset of decompensation in healthy volunteers?
Observational (n=101)
Does compensatory reserve measurement (CRM) provide greater sensitivity and specificity than heart rate variability (HRV) and heart rate complexity (HRC) to detect progressive reductions in central circulating blood volume and onset of decompensation in healthy volunteers?
Effect estimate: ROC AUC 0.93
Absolute Event Rate: 0.93% vs 0.86%
Compensatory reserve measurement provides superior sensitivity and specificity compared to heart rate variability and complexity for detecting progressive reductions in central blood volume and predicting hemodynamic decompensation.
CRM may aid early hypovolemia detection in controlled settings; leaves open utility in trauma patients pending prospective validation.
BACKGROUND: Hemorrhage remains the leading cause of death following traumatic injury in both civilian and military settings. Heart rate variability (HRV) and heart rate complexity (HRC) have been proposed as potential "new vital signs" for monitoring trauma patients; however, the added benefit of HRV or HRC for decision support remains unclear. Another new paradigm, the compensatory reserve measurement (CRM), represents the integration of all cardiopulmonary mechanisms responsible for compensation during relative blood loss and was developed to identify current physiologic status by estimating the progression toward hemodynamic decompensation. In the present study, we hypothesized that CRM would provide greater sensitivity and specificity to detect progressive reductions in central circulating blood volume and onset of decompensation as compared with measurements of HRV and HRC. METHODS: Continuous, noninvasive measurements of compensatory reserve and electrocardiogram signals were made on 101 healthy volunteers during lower-body negative pressure (LBNP) to the point of decompensation. Measures of HRV and HRC were taken from electrocardiogram signal data. RESULTS: Compensatory reserve measurement demonstrated a superior sensitivity and specificity (receiver operator characteristic area under the curve [ROC AUC] = 0.93) compared with all HRV measures (ROC AUC ≤ 0.84) and all HRC measures (ROC AUC ≤ 0.86). Sensitivity and specificity values at the ROC optimal thresholds were greater for CRM (sensitivity = 0.84; specificity = 0.84) than HRV (sensitivity, ≤0.78; specificity, ≤0.77), and HRC (sensitivity, ≤0.79; specificity, ≤0.77). With standardized values across all levels of LBNP, CRM had a steeper decline, less variability, and explained a greater proportion of the variation in the data than both HRV and HRC during progressive hypovolemia. CONCLUSION: These findings add to the growing body of literature describing the advantages of CRM for detecting reductions in central blood volume. Most importantly, these results provide further support for the potential use of CRM in the triage and monitoring of patients at highest risk for the onset of shock following blood loss.
No takes yet. Share an insight, caveat, or question.
Schlotman et al. (2020) conducted an observational in Progressive hypovolemia (simulated) (n=101). Compensatory reserve measurement (CRM) vs. Heart rate variability (HRV) and heart rate complexity (HRC) was evaluated on Detection of progressive reductions in central circulating blood volume and onset of decompensation (ROC AUC 0.93). Compensatory reserve measurement demonstrated superior accuracy (ROC AUC 0.93) compared with heart rate variability (AUC ≤0.84) and complexity (AUC ≤0.86) for detecting hemodynamic decompensation.