Key result
Mimicked neo-sinus on a PCU aortic valve increases in-vitro clot mass ~21% versus no neo-sinus.
Why the study?
The neo-sinus created after TAVR increases the thrombogenicity of transcatheter heart valves, but testing methods to evaluate this thrombogenicity in vitro are lacking.
Does the presence of a neo-sinus increase thrombogenicity in an in-vitro model of transcatheter aortic valve replacement?
Does the presence of a neo-sinus increase thrombogenicity in an in-vitro model of transcatheter aortic valve replacement?
Effect estimate: 21.21% higher mass
Absolute Event Rate: 0.064% vs 0.0528%
In an in-vitro model, the presence of a neo-sinus mimicking post-TAVR anatomy altered flow dynamics and increased clot formation compared to a valve without a neo-sinus, supporting clinical observations of subclinical leaflet thrombosis.
In-vitro clot increase with neo-sinus suggests design caution; leaves open clinical thrombosis risk.
PURPOSE: Transcatheter aortic valve replacement (TAVR) is the standard treatment for patients with aortic diseases at high surgical risk. Transcatheter heart valve prostheses (THV) are inserted into the aortic valve, creating a new area between the native and artificial leaflets. This area, known as neo-sinus, increases the thrombogenicity of THVs. But there is a lack of testing methods that evaluate thrombogenicity in vitro. METHODS: To analyze the flow field within the native sinus and the neo-sinus, Particle Image Velocimetry (PIV) was performed with a thrombosis tester. Additionally, a comparative study was conducted with porcine blood on two polycarbonate urethane valves, with and without neo-sinus, respectively. Blood samples collected every hour were analyzed for platelet count, coagulation via ROTEM parameters, and plasma-free hemoglobin. Thrombus formation was detected optically. RESULTS: The PIV measurements yield a physiological flow field in the aortic root that were consistent with those reported in literature. The analyzed blood parameters reveal no obvious difference between the valve with neo-sinus and the valve without. A higher amount of thrombus material for the valve with neo-sinus was found. CONCLUSION: The visualized flow field shows low velocities and stagnation zones due to the presence of native leaflets. Clot formation at the heart valve prostheses are in accordance with in-vivo findings. The benchmark of the two valves indicates an increased thrombogenic potential due to the neo-sinus. The thrombosis tester simulates the natural environment after TAVR. Thereby, newly developed THVs can be evaluated in vitro and consequently optimized regarding their thrombogenicity.
No takes yet. Share an insight, caveat, or question.
Thoenissen et al. (2025) studied Aortic valve replacement (TAVR). Valve with neo-sinus vs. Valve without neo-sinus was evaluated on Total mass of clots (21.21% higher mass). The presence of a mimicked neo-sinus on a polycarbonate urethane aortic valve resulted in a 21.21% higher mass of clots compared to a valve without a neo-sinus in an in-vitro blood test.