Key result
ACURATE neo achieves comparable device success to Sapien 3 in transfemoral TAVI.
Why the study?
Differences in clinical outcomes between the balloon-expandable Sapien 3 and self-expanding ACURATE neo valve systems in a real-world transfemoral TAVI population were unclear.
Does the ACURATE neo valve yield comparable device success and early clinical safety compared to the Sapien 3 valve in patients with severe aortic stenosis undergoing transfemoral TAVI?
Cohort (n=345)
Yes
Does the ACURATE neo valve yield comparable device success and early clinical safety compared to the Sapien 3 valve in patients with severe aortic stenosis undergoing transfemoral TAVI?
The self-expanding ACURATE neo and balloon-expandable Sapien 3 valves offer comparable, excellent device success and short-term clinical safety in real-world patients undergoing transfemoral TAVI.
Answers are grounded in this paper and cite their sources.
Comparable device success supports either valve in real-world TAVI; leaves open need for randomized confirmation of equivalence.
OBJECTIVES: To compare the outcomes of transfemoral ACURATE neo (NEO) and Sapien 3 (S3) patients in terms of device success and clinical safety outcomes using a propensity score analysis. BACKGROUND: Differences in clinical outcomes between the latest-generation balloon-expandable S3 and self-expanding NEO in a "real-world transfemoral TAVI population" are still unclear. METHODS: We compared up to 6 months clinical outcomes using a propensity score analysis (inverse probability of treatment weighting [IPTW]) to account for differences in baseline characteristics. RESULTS: A total of 345 patients underwent transfemoral transcatheter aortic valve implantation (TAVI) with either NEO or S3 at two centers in the Netherlands. Composite device success and early safety endpoints were comparable between NEO and S3 (Device success: IPTW-adjusted OR: 0.35 [95% CI: 0.12-1.18], and early safety: IPTW-adjusted OR: 0.51 [95% CI: 0.19-1.38]). Six-months mortality was 5.3 versus 3.6%, stroke was 2.8 versus 3.3%, and pacemaker rate was 6.1 versus 8.6%, respectively with p = NS. Mean aortic gradient was lower in the NEO group (5.72 ± 2.47 vs. 9.05 ± 3.48; p = <.001), with a comparable rate of moderate or severe paravalvular leak (0 versus 2.1%; p = NS). CONCLUSIONS: Device success and clinical safety outcomes were comparable for both valves. Up to 6-months follow-up clinical outcomes and mortality rate remained excellent. Mean aortic gradient was lower after ACURATE neo implantation.
No takes yet. Share an insight, caveat, or question.
Kooistra et al. (2020) conducted a cohort in Symptomatic severe aortic stenosis (n=345). ACURATE neo valve vs. Sapien 3 valve was evaluated on Composite device success (IPTW-adjusted OR 0.35, 95% CI 0.12-1.18, p=0.061). The ACURATE neo valve demonstrated comparable composite device success to the Sapien 3 valve in patients undergoing transfemoral TAVI.