Key result
Improved EEMD with FMCW radar accurately estimates heart rate within ~4 bpm.
Why the study?
In non-contact vital sign detection using FMCW radar, standard EEMD requires improvement to overcome clutter and noise characteristics in public places for separating respiratory and heartbeat signals.
An improved signal processing algorithm for FMCW radar enables accurate, non-contact detection of heart rate with errors below 4 bpm.
May support contactless monitoring in clinical settings; extends prior algorithms but leaves open prospective validation.
Non-contact vital sign detection technology has brought a more comfortable experience to the detection process of human respiratory and heartbeat signals. Ensemble empirical mode decomposition (EEMD) is a noise-assisted adaptive data analysis method which can be used to decompose the echo data of frequency modulated continuous wave (FMCW) radar and extract the heartbeat and respiratory signals. The key of EEMD is to add Gaussian white noise into the signal to overcome the mode aliasing problem caused by original empirical mode decomposition (EMD). Based on the characteristics of clutter and noise distribution in public places, this paper proposed a static clutter filtering method for eliminating ambient clutter and an improved EEMD method based on stable alpha noise distribution. The symmetrical alpha stable distribution is used to replace Gaussian distribution, and the improved EEMD is used for the separation of respiratory and heartbeat signals. The experimental results show that the static clutter filtering technology can effectively filter the surrounding static clutter and highlight the periodic moving targets. Within the detection range of 0.5 m~2.5 m, the improved EEMD method can better distinguish the heartbeat, respiration, and their harmonics, and accurately estimate the heart rate.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2022) studied Healthy subjects (Algorithm validation) (n=6). Improved Ensemble Empirical Mode Decomposition (EEMD) algorithm with static clutter filtering vs. Traditional EMD/EEMD algorithms and MI3 contact smart wristband was evaluated on Heart rate estimation accuracy (root mean square error compared to contact sensor). The improved EEMD algorithm with FMCW radar effectively distinguished heartbeat and respiration components, accurately estimating heart rate with a root mean square error of less than 4 bpm.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: