Demonstrates nontargeted calibration approximates uptake rates in diverse compounds, suggesting a new analytical approach.
Conventional calibration of passive air samplers (PASs) is limited by the accessibility of analytical standards required for quantifying mass in targeted analysis. Here, we introduce a method using relative peak areas measured with nontargeted analysis to approximate PAS uptake rates for diverse features. We tested the approach with mesh-sheltered polydimethylsiloxane (PDMS) foam and glass fiber filter (GFF) PASs. Generic uptake rates were 1.16 ± 0.34 m 3 day –1 dm –2 for 67 features measured in PDMS foam and 0.50 ± 0.33 m 3 day –1 dm –2 for 48 features measured in GFF, comparable to uptake rates from the targeted calibration literature. The generic uptake rate for PDMS-Mesh showed low variability (relative standard deviation─RSD = 29%) with no relationship to predicted physical–chemical properties. The generic rate for GFF-Mesh was instead 0.61 ± 0.20 m 3 day –1 dm –2 when including only features with predicted log octanol-air partition ratios (K OA s) < 8.5 log units, showing decreased variability (from 66% to 32% RSD). Specific uptake rates should be used for features with K OA s > 8.5 log units measured on GFF-Mesh. Our results demonstrate that nontargeted calibration can characterize PAS uptake for diverse compounds comparably to targeted calibration, which will expand our understanding of uptake kinetics without requiring analytical standards.