June 23, 2026Journal of Biomedical Materials Research Part B Applied Biomaterials

Strut Thickness and Species‐Specific Healing Are Key Considerations in Developing Zinc‐Based Biodegradable Arterial Stents

View Full Paper
Ask AI
Bookmark
Share

Authors

LMLea M. MorathSRShebeer A. RahimCBCole A. Baker

Discussion

Loading...

Member takes

Overview

Randomized trial examines healing responses in biodegradable zinc stents, indicating model selection is crucial.

Key Points

  • This research aims to evaluate the performance and healing responses of different zinc-based biodegradable stents.
  • Examined three zinc-based alloys: AMZ, ACMZ, and CMM, focusing on strut thickness and processing.
  • Conducted in vivo implantation in abdominal aortas of mice and rats to analyze immune responses.
  • Assessed thrombogenicity and hemocompatibility through in vitro assays and ex vivo analyses.
  • Mice showed significant inflammation and necrosis, indicating poor biocompatibility (p<0.05).
  • Rats had minimal inflammation and reduced neointimal growth with 0.12 mm struts compared to 0.25 mm implants.
  • In vitro assays demonstrated low thrombogenicity with lower fibrin generation (p<0.01 compared to controls).