Key result
Sirolimus-eluting bioabsorbable scaffolds reduce renal artery inflammation ~31% versus unloaded scaffolds in pigs.
Why the study?
Does an absorbable sirolimus-eluting polymer scaffold improve biocompatibility and reduce inflammation compared to unloaded scaffolds or bare metal stents in porcine renal arteries?
Does an absorbable sirolimus-eluting polymer scaffold improve biocompatibility and reduce inflammation compared to unloaded scaffolds or bare metal stents in porcine renal arteries?
Answers are grounded in this paper and cite their sources.
Absorbable sirolimus-eluting polymer scaffolds demonstrate mechanical stability and reduced inflammatory response compared to unloaded scaffolds in a porcine renal artery model.
BACKGROUND: This study aims to assess the technical feasibility and biocompatibility of interventional application of a fully absorbable stent in renal arteries (RAs) of healthy pigs. METHODS: Absorbable poly(L-lactide) (PLLA)/poly(4-hydroxybutyrate) scaffolds either loaded with sirolimus (SIR) coating or unloaded and permanent bare metal stents (BMSs) served as control were implanted interventionally into RAs of 16 female pigs via the left common iliac artery (8F-sheath). Peroral dual antiplatelet therapy was administered throughout the study. Stented RA segments were explanted after 4 weeks and assessed histomorphometrically. RESULTS: and 23.0 ± 4.5%). Incorporation of SIR reduced the significantly higher inflammation of unloaded scaffolds; however, not to a level compared to BMS (unloaded PLLA-based: 1.3 ± 0.3; SIR-PLLA-based: 0.9 ± 0.3; BMS: 0.5 ± 0.1). In contrast, BMS showed a slightly elevated vascular injury score (0.8 ± 0.1) compared to the unloaded PLLA-based (0.6 ± 0.2) and the SIR-PLLA-based (0.4 ± 0.04) groups. CONCLUSION: Absorbable scaffolds showed sufficient mechanical stability after porcine RA stenting. By incorporation of SIR, a significant reduction of the inflammatory and neointimal response to the unloaded polymer scaffolds was seen without systemic toxicity or thrombotic complications. However, the greater strut height of the polymer scaffold is a major limitation and need to be addressed in future work on the stent design.
No takes yet. Share an insight, caveat, or question.
Kischkel et al. (2025) studied healthy pigs (n=16). Absorbable sirolimus-eluting polymer scaffold vs. Unloaded polymer scaffolds and permanent bare metal stents was evaluated on Inflammation score. Sirolimus-eluting absorbable scaffolds reduced inflammation versus unloaded scaffolds but had higher inflammation than bare metal stents in porcine renal arteries.
Synapse has enriched 10 related papers. Consider them for comparative context: