Key result
Common HbA1c devices showed optimal incremental reclassification rates for diabetes diagnosis (0.11-0.22%), but the Abbott Afinion failed the minimal standard for prediabetes diagnosis (8.62%).
A novel simulation method based on incremental reclassification rates can help external quality assurance programs identify poorly performing in vitro diagnostic devices.
Background External quality assurance (EQA) programs must assess whether the analytical performance of commonly used in vitro diagnostic (IVD) devices is fit for their intended clinical use. We developed a method for this assessment based on reclassification of results at clinical decision limits and applied it to results from a haemoglobin A 1c (HbA 1c ) EQA program. Content Using a commutable EQA program to provide analytical performance characteristics, we simulated 10,000 measurements of the HbA 1c value of each of 6,326 subjects in the National Health and Nutrition Survey (NHANES) by common IVD devices and the NHANES device. A parametric empirical Bayes approach was used to account for regression to the mean. We estimated the rates at which the IVD devices would reclassify results and compared them with the estimated reclassification rate for the NHANES device, calculating the difference between them as the ‘incremental reclassification rate’ (IRR). IRRs were ‘optimal’ if <1.5 %, ‘desirable’ if <3.0 %, and ‘minimal’ if <4.7 %. Summary and outlooks EQA performance data were available for five HbA 1c devices: Abbott Afinion, Bio-Rad D series, Roche Cobas, Sebia Capillarys, and Siemens DCA. The median IRRs were optimal for all methods for diabetes diagnosis (0.11–0.22 %) and for diabetes monitoring (−0.51–1.01 %), with negative IRRs representing lower reclassification rates than the NHANES device. For prediabetes diagnosis, the Afinion failed the minimal standard (8.62 %), but median IRRs were desirable for the other methods (1.43–2.69 %). EQA providers should consider reporting IRR for suitable analytes to help identify poorly performing IVD devices.
No takes yet. Share an insight, caveat, or question.
Farrell et al. (2026) studied Diabetes and prediabetes (n=6,326). Common in vitro diagnostic (IVD) devices for HbA1c vs. NHANES device was evaluated on Incremental reclassification rate (IRR). Common HbA1c devices showed optimal incremental reclassification rates for diabetes diagnosis (0.11-0.22%), but the Abbott Afinion failed the minimal standard for prediabetes diagnosis (8.62%).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: