Hospital laboratory HbA1c methods in Sweden improved over 20 years to fulfill analytical performance specifications, whereas point-of-care test accuracy varied over time.
Observational
Yes
Small variations in the trueness of HbA1c testing methods can significantly impact the mean HbA1c values of patient cohorts, highlighting the need for continuous external quality assessment.
Point-of-care HbA1c accuracy requires continued scrutiny; leaves open optimal standardization strategies across settings.
BACKGROUND: The accuracy and trueness of results from a laboratory test, such as the HbA1c test, should not be taken for granted but must be checked continuously. A tool for this is the participation in external quality assessment (EQA) for all laboratories performing the HbA1c-test. An additional possibility to detect changes in trueness is to monitor variations in patient cohort mean or median values that is not explained by changes in treatment or selection of patients. METHODS: Results reported to an EQA scheme for HbA1c during 20 years have been extracted from Equalis database. The results are compared to current analytical performance specifications (APS) and to the mean HbA1c levels for the Swedish population of persons with type 2 diabetes. RESULTS: The accuracy of the HbA1c test has improved during the period. The hospital lab methods used in Sweden now fulfil APS agreed by professional organizations in Sweden. The accuracy for point-of-care tests (POCT) methods vary over time and fulfil APS for some periods. The bias found for some of the methods might explain changes seen in patient mean values for HbA1c in Sweden during the period 2007-2017. CONCLUSIONS: The global standardization of HbA1c has resulted in an improved comparability for HbA1c-results worldwide. But even small variation in trueness for the methods in use might have important impact on mean HbA1c values for cohorts of patients. When a systematic error is observed for a specific method it is therefore essential that manufacturers correct the method without delay.
No takes yet. Share an insight, caveat, or question.
Gunnar Nordin (2018) conducted an observational in Type 2 diabetes. HbA1c test methods vs. Analytical performance specifications (APS) was evaluated on Accuracy and trueness of HbA1c test results. Hospital laboratory HbA1c methods in Sweden improved over 20 years to fulfill analytical performance specifications, whereas point-of-care test accuracy varied over time.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: