Measures beam quality factors for proton beams, validating standards for clinical use.
Purpose This study aimed to experimentally determine beam quality correction factors ( k Q ) for NE2571 and PTW 30013 chambers in clinical proton beams using water calorimetry and to compare the results with the TRS-398 rev.1 (2024). Methods Measurements were performed in the middle of the spread-out Bragg peak of a 150 MeV double-scattered proton beam ( R res = 7.5 g cm −2 ). k Q values were derived by comparing chamber readings with the absorbed dose to water determined by a sealed-water calorimeter operating at 4 ∘C. Results The determined k Q values were 1.015 ± 0.009 (NE2571) and 1.012 ± 0.009 (PTW 30013). The NE2571 chamber showed excellent agreement with TRS-398 rev.1 (2024), with a deviation of only 0.7 %. The PTW 30013 exhibited a larger deviation (1.3 %), although within the standard uncertainty of the protocol. Derived ( p ch / p dis ) Q values were consistent with scattered beam literature but showed systematic differences compared to for scanned beams. Conclusions Results validate TRS-398 rev.1 (2024) for the NE2571 under scattered beam conditions. However, the PTW 30013 deviation and observed trends in ( p ch / p dis ) Q warrant further investigation into chamber-specific responses and the influence of beam delivery modalities.