Demonstrates quantitative monitoring of o-phthalaldehyde disinfectants, suggesting improved accuracy in endoscope reprocessing.
Routine monitoring of ortho-phthalaldehyde (OPA) disinfectants is critical for endoscope reprocessing, yet commercial test strips suffer from subjective visual ambiguity, strict manual timing, and susceptibility to sample matrix dilution. This study proposes a portable multi-spectral colorimetric sensing platform paired with structurally engineered microfluidic plastic strips for quantitative OPA monitoring. The strips utilize a confined microfluidic geometry to achieve capillary-driven volumetric self-metering (5.4 μL), while cross-hatched micro-structures eliminate edge pooling, yielding uniform colorimetric responses. Analytically, the system integrates a matrix-matched reagent formulation, an interference-free indicator, and an automated steady-state ratiometric readout algorithm to counteract physical dilution and spectral interference. Cross-validation against a capillary electrophoresis benchmark confirmed quantitative accuracy (R2 = 0.9684) under physical dilution of real-world CIDEX OPA solutions. This correlation facilitated a matrix-compensated 0.32% diagnostic threshold for unambiguous, automated “[PASS]” or “[FAIL]” alerts. Ultimately, this scalable, cost-effective microfluidic architecture provides an objective point-of-care diagnostic solution, demonstrating translational potential for broad dry chemistry optical detection.