Key result
Jellyfish Method resolves initial pentaspline PFA failure to achieve 100% successful PVI.
Why the study?
In pentaspline PFA for atrial fibrillation, anatomical challenges in small pulmonary veins frequently cause unstable contact and high-impedance errors, leading to procedural failure.
Does the Jellyfish Method improve lesion uniformity and procedural success in anatomically challenging pulmonary veins during pentaspline PFA?
Cohort (n=149)
Does the Jellyfish Method improve lesion uniformity and procedural success in anatomically challenging pulmonary veins during pentaspline PFA?
The Jellyfish Method is an effective troubleshooting technique for pentaspline PFA in small pulmonary veins, improving lesion uniformity and procedural success.
Answers are grounded in this paper and cite their sources.
May facilitate PVI troubleshooting in select cases; hypothesis-generating and requires prospective validation before practice change.
Background/Introduction Pulsed field ablation (PFA) with the pentaspline system is effective for atrial fibrillation, but anatomical challenges, particularly in small pulmonary veins (PVs) like the right inferior PV (RIPV), frequently cause unstable contact and high-impedance errors, leading to procedural failure. Purpose We introduce a novel troubleshooting technique, the "Jellyfish Method," designed to overcome these anatomical limitations. We aimed to validate its biophysical advantages in-vitro and report its prevalence and utility in a consecutive clinical cohort. Methods The "Jellyfish Method" is a novel configuration where the five splines are aligned parallel to the PV's course, optimising contact. First, to characterise its biophysical properties, an in-vitro study using 17-mm and 20-mm potato models compared lesion depth and uniformity (minimum depth) between Jellyfish, Basket, and Olive configurations. Second, we retrospectively analysed 149 consecutive patients undergoing first-time PVI to determine the clinical prevalence, anatomical distribution, and success of this method. Results The in-vitro study confirmed the method's biophysical superiority for small anatomies. In the 17-mm model, the Jellyfish configuration created significantly deeper and, crucially, more uniform lesions. It achieved a minimum perpendicular depth of 2.7 mm, far superior to the Basket configuration (0.5 mm), which showed gaps. This in-vitro advantage was directly reflected in clinical practice. The Jellyfish Method was required as a troubleshooting technique in 34 of 149 patients (22.8%) to overcome impedance errors or instability. The technique was most frequently deployed in the Right Inferior PV (32/34 patients, 94.1%), followed by the Right Superior PV (9/34, 26.5%). In all 34 cases, switching to the Jellyfish method immediately resolved the initial failure, enabling successful PVI. Conclusion(s) The "Jellyfish Method," validated in-vitro for superior lesion uniformity in small anatomies, is a critical troubleshooting technique required in over 20% of pentaspline PFA procedures. It provides a simple, reproducible solution for anatomically challenging PVs, particularly the RIPV, ensuring procedural success by preventing impedance errors and expanding the system's applicability.
No takes yet. Share an insight, caveat, or question.
Kanda et al. (2026) conducted a cohort in atrial fibrillation (n=149). Jellyfish Method (pentaspline PFA configuration) vs. Basket configuration (in-vitro) was evaluated on Successful PVI after initial failure. The Jellyfish Method successfully resolved initial procedural failure and enabled successful pulmonary vein isolation in all 34 patients requiring it.
Synapse has enriched 10 related papers. Consider them for comparative context: