Key result
GUIDE algorithm predicts ECGI-defined ventricular resynchronization during LBBAP with ~72% positive predictive value.
Why the study?
Existing ECG criteria to confirm conduction system capture during LBBAP are complex, and previous validation of the simplified GUIDE algorithm was retrospective and limited to the final lead position.
Does the GUIDE algorithm accurately predict ventricular resynchronization during left bundle branch area pacing?
Observational (n=52)
Does the GUIDE algorithm accurately predict ventricular resynchronization during left bundle branch area pacing?
The GUIDE algorithm offers high specificity but limited sensitivity for predicting ventricular resynchronization during LBBAP, suggesting ECGI may provide complementary real-time assessment.
Answers are grounded in this paper and cite their sources.
GUIDE's modest accuracy for LBBAP resynchronization warrants clinical caution; leaves open ECGI's complementary role in prospective trials.
Background Several electrocardiographic (ECG) criteria have been proposed to confirm conduction system capture during left bundle branch area pacing (LBBAP), but their application remains complex. We previously reported a stepwise ECG algorithm (GUIDE) to simplify implantation and ensure ventricular resynchronization. Its validation was retrospective and limited to final lead position [1]. Objective To prospectively evaluate the diagnostic performance of GUIDE algorithm to predict ventricular resynchronization and to analyze the added value of electrocardiographic imaging (ECGI), during LBBAP implantation. Methods Fifty-two patients undergoing LBBAP were included. ECGI ventricular activation maps were obtained as the lead was progressively advanced through the interventricular septum. Resynchronization was defined as a reduction of the left ventricular activation time (LVAT95) > 15ms. ECGs were assessed at each screwing stage and classified as GUIDE+ or GUIDE– (Figure 1). Results A total of 346 lead positions were analyzed (52 baseline, 194 GUIDE–, 100 GUIDE+). Progressive screwing reduced LVAT95: 76 ms at baseline, 65 ms in GUIDE– and 54 ms in GUIDE+ (p < 0.0001). The proportion of resynchronized positions increased from 49% in GUIDE– to 72% in GUIDE+ (p < 0.0001). The GUIDE algorithm showed 43% sensitivity, 78% specificity, 72% positive and 50% negative predictive value for ECGI-defined resynchronization. Conclusions GUIDE algorithm demonstrates high specificity but limited sensitivity in predicting ventricular resynchronization during LBBAP implantation. While GUIDE+ criteria indicate optimal resynchronization, some patients achieve resynchronization despite not meeting them. ECGI provides real-time assessment of ventricular activation and may complement decision-making in the clinical routine.
No takes yet. Share an insight, caveat, or question.
Regany et al. (2026) conducted an observational in Patients undergoing left bundle branch area pacing (LBBAP) (n=52). GUIDE algorithm vs. GUIDE- criteria was evaluated on Ventricular resynchronization (reduction of left ventricular activation time > 15ms) (p=< 0.0001). The GUIDE algorithm predicted ECGI-defined ventricular resynchronization during left bundle branch area pacing with 43% sensitivity, 78% specificity, and a 72% positive predictive value.