Key result
Implantation of the MedEng-BIO xenopericardial bioprosthesis decreased the peak pressure gradient by 57.8% and the mean gradient by 51.9% at two years post-surgery.
Why the study?
The study was conducted to assess midterm results after implantation of a frame-mounted xenopericardial bioprosthesis into the aortic position using the easy-change system.
Does implantation of a frame-mounted xenopericardial bioprosthesis improve echocardiographic hemodynamic parameters in patients requiring aortic valve replacement?
Observational (n=166)
No
Does implantation of a frame-mounted xenopericardial bioprosthesis improve echocardiographic hemodynamic parameters in patients requiring aortic valve replacement?
The MedEng-BIO frame xenopericardial bioprosthesis provides satisfactory midterm hemodynamic correction for patients undergoing aortic valve replacement.
Observed gradient reductions suggest favorable hemodynamics; leaves open need for randomized comparisons of durability and outcomes.
Purpose . The purpose of this work was to study the midterm results after the implantation of a frame-mounted xenopericardial bioprosthesis into the aortic position with the easy-change system. Material and Methods . A total of 166 patients with mean age of 69.5 years underwent aortic valve replacement by bioprosthesis with the easy-change system at Cardiology Research Institute, Tomsk National Research Medical Center, Russian Academy of Sciences between October, 2016 and October, 2020. Preoperative aortic valve peak pressure gradient was 79.6 mmHg; the mean gradient was elevated up to 46.3 mmHg. Echocardiography studies of aortic prosthesis and the left ventricle (LV) were performed in all patients by Vivid 7, GE, and IE 33 systems (Philips) preoperatively, before the discharge (day 14 after surgery on average), and 6–12 months after surgery. Results . The control echocardiography studies, performed 6–12 months after surgery, showed that the hemodynamic characteristics of MedEng-BIO aortal bioprostheses were satisfactory. No statistically significant changes in the LV function were detected. The effective orifice areas of MedEng-BIO valves were 0.87 cm 2 for the valve size 21, 1.1 cm 2 for size 23, and 1.47 cm 2 for size 25 mm. The peak pressure gradient decreased by 57.8% and the mean gradient decreased by 51.9% two years after surgery. Conclusion . Obtained data suggest that the implantation of MedEng-BIO aortal xenopericardial bioprosthesis into the aortic position lead to adequate correction of hemodynamic parameters.
No takes yet. Share an insight, caveat, or question.
Петлин et al. (2021) conducted an observational in Aortic valve disease requiring replacement (n=166). MedEng-BIO aortal xenopericardial bioprosthesis with easy-change system was evaluated on Hemodynamic characteristics including peak and mean pressure gradients. Implantation of the MedEng-BIO xenopericardial bioprosthesis decreased the peak pressure gradient by 57.8% and the mean gradient by 51.9% at two years post-surgery.