BACKGROUND: Multi-frequency bioelectrical impedance analysis (MF-BIA) offers enhanced body composition outcomes in a time-efficient manner. The accuracy of stand-up MF-BIA compared against a four-compartment (4C) criterion lacks evidence. OBJECTIVES: To validate a stand-up MF-BIA compared to a 4C criterion for fat mass (FM), fat-free mass (FFM) and body fat percentage (%fat). SUBJECTS/METHODS: ) young adults were measured for body composition determined from a stand-up MF-BIA and 4C model. Validity statistics included total error (TE) and standard error of the estimate (SEE) to examine prediction error between methods. RESULTS: For the total sample, prediction error was the highest for %fat (TE = 4.2%; SEE = 3.9%) followed by FM (TE = 2.4 kg; SEE = 2.2 kg) and FFM (TE = 2.4 kg; SEE = 2.2 kg). In men, %fat (TE = 2.5%; SEE = 2.2%) and FM (TE = 1.9 kg; SEE = 1.6 kg) were ideal; FFM was similar to FM (TE = 1.9 kg; SEE = 1.6 kg). In women, %fat (TE = 4.7%; SEE = 4.4%) ranged from good to fairly good, and FM was very good to excellent (TE = 2.6 kg; SEE = 2.4 kg); FFM was similar to FM (TE = 2.6 kg; SEE = 2.3 kg). CONCLUSIONS: Stand-up MF-BIA may overestimate %fat and FM, and underestimate FFM compared to a 4C model. FM and FFM estimates from MF-BIA demonstrate good agreement to a 4C model and may be a practical measure of body composition in normal-weight adults. The highest error was seen in %fat for both sexes, with greater error in women.