Evaluates noise optimization of capacitance multipliers in biomedical signal processing, suggesting effectiveness in low-frequency applications.
This study presents a detailed evaluation of two capacitance multiplier configurations with a focus on noise optimization in second order low pass filters. The proposed multipliers are based on MO-VDTA (Multi Output Voltage Differencing Transconductance Amplifier) and MO-CDTA (Multi Output Current Differencing Transconductance Amplifier) architectures. In the MO-VDTA based design, the multiplication factor ranges from 120 to 750 (CB= 1pF), controlled by the bias current IB. The designed filter provides a tunable cut-off frequency ranging from 2 kHz to 12.4 kHz (CB= 1pF). In additon, the MO-CDTA based structure enables the multiplication factor ‘k’ to be adjusted between 400 and 1250 (CB= 2pF) by varying the gate-source voltage (VGS) of an external PMOS transistor connected to the Z terminal. This configuration is used in a low pass filter with a tunable cut-off frequency between 23.6 kHz and 91 kHz (CB= 2pF), suitable for ultrasound applications. Additionally, this study includes a comprehensive noise analysis of both capacitance multiplier based filters, demonstrating their effectiveness in low frequency and biomedical signal processing environments.
No takes yet. Share an insight, caveat, or question.
Sakacı et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: