Key result
Same-access snaring retrieves a ruptured TAVI balloon through an ipsilateral sheath without vascular injury.
Why the study?
Balloon rupture with distal fragment retention during balloon-expandable TAVI is a rare, technically demanding complication lacking a standardized retrieval protocol.
Case Report (n=1)
No
Demonstrates a stepwise, single-access percutaneous snare technique to safely retrieve a ruptured and retained TAVI balloon fragment via the existing ipsilateral sheath after correcting sheath tip deformation, avoiding surgical conversion or contralateral large-bore access.
May facilitate percutaneous bailout in rare TAVI balloon rupture; leaves open prospective validation before wider adoption.
BACKGROUND: Balloon-expandable valves are among the most commonly used platforms for transfemoral transcatheter aortic valve implantation (TAVI). Although deployment-related complications are uncommon, balloon rupture with distal fragment retention is a rare but technically demanding scenario with no standardized retrieval protocol. CASE SUMMARY: A 79-year-old patient underwent a transfemoral TAVI with a balloon-expandable valve. Balloon rupture during deployment resulted in partial retention of the distal balloon segment. The patient remained hemodynamically stable without annular or aortic injury. The retained fragment was successfully retrieved through the same femoral access site using a snare-based technique, without contralateral large-bore access or surgical conversion. DISCUSSION: Same-access snaring through the ipsilateral E-sheath-after correcting the deformed tip to restore coaxial alignment-allows safe, controlled extraction of the retained balloon fragments through a single-access site. Recognizing and correcting sheath tip deformation before snare advancement is a critical technical prerequisite. TAKE-HOME MESSAGES: Balloon rupture during balloon-expandable TAVI may result in retained balloon fragments. Same-access snaring through the ipsilateral E-sheath is a feasible and safe first-line strategy. Sheath deformation must be recognized and corrected to ensure successful extraction without escalation to additional access or surgery.
No takes yet. Share an insight, caveat, or question.
Maaroufi et al. (2026) conducted a case report in Low-flow low-gradient aortic stenosis undergoing transfemoral balloon-expandable TAVI complicated by balloon rupture and distal fragment retention (n=1). Same-access snare-based retrieval through the ipsilateral 16-F E-sheath after sheath tip straightening with a 0.035-inch J-tip guidewire was evaluated on Successful extraction of retained distal balloon fragment and nosecone without vascular injury or surgical conversion. Same-access snaring through a straightened ipsilateral 16-F E-sheath successfully retrieved a retained ruptured TAVI balloon fragment in 1 patient without vascular injury or surgical conversion.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: