Key result
Repeat balloon aortic valvuloplasty successfully frees an infolded, entrapped self-expandable TAVR delivery system.
Why the study?
Valve infolding is a rare but serious complication after implantation of self-expanding transcatheter aortic valve prostheses that can cause difficulty in device removal.
Case Report (n=1)
May offer bailout for rare TAVR infolding complications; hypothesis-generating and leaves open need for broader validation.
Transcatheter aortic valve replacement (TAVR) is a well-established treatment for severe aortic stenosis. Advances in the devices used and operators' technique have reduced the frequency of complications. However, valve infolding is a rare but serious outcome after the implantation of self-expanding prostheses. We report a case of a successful bailout of a device that was difficult to remove because of valve infolding. TAVR using a 26 mm Evolut PRO+ system (Medtronic) was planned for a 91-year-old woman with severe aortic stenosis. After the valve was deployed in a satisfactory position on the second release, the system could not be removed because the nose cone was hooked to the basal frame of the deployed valve. To overcome this situation, an additional balloon was inserted from the contralateral femoral side and inflated, and we extracted the system successfully by pulling out the device while simultaneously deflating the balloon. Postoperative computed tomography revealed valve infolding, which was considered to cause the difficulty in system removal. Infoldings of self-expandable prostheses should be considered when faced with difficulty in removing the catheter system, and the method elucidated in this case report can be effective to manage it.
No takes yet. Share an insight, caveat, or question.
Yashige et al. (2022) conducted a case report in Severe aortic stenosis (n=1). Additional balloon aortic valvuloplasty was evaluated. Additional balloon aortic valvuloplasty successfully facilitated the extraction of a hooked self-expandable TAVR catheter system caused by valve infolding.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: