Key result
TAVI prosthetic valve infolding occurs predominantly with ≥29-mm valves, linked to ~94% of reported cases.
Why the study?
Prosthetic valve infolding is a rare but severe complication of TAVI with self-expanding valves, and currently available clinical data are limited and fragmented.
Systematic Review (n=34)
Yes
Alerts clinicians to rare infolding risk with self-expanding TAVI valves; leaves open optimal prevention and management strategies.
BACKGROUND: Prosthetic valve infolding is a rare but severe complication of transcatheter aortic valve implantation (TAVI) with self-expanding valves. However, currently available clinical data are limited and fragmented. OBJECTIVES: This report aims to provide a comprehensive overview of this complication focusing on predisposing factors, clinical presentation, diagnostic findings, treatment and clinical outcomes. METHODS: A systematic review of the literature was performed to identify cases of infolding occurring during TAVI with self-expanding valves published until August 2020. These data were pooled with all the retrospectively identified infolding cases occurred at San Raffaele Scientific Institute between December 2014 and August 2020. RESULTS: A total of 34 cases were included. Among patients with available data, 38% received a first-generation CoreValve, and 62% a second-generation Evolut R (82%) or Evolut PRO (18%). Infolding occurred mostly with ≥29-mm valves (94%). Predisposing factors included resheathing of a second-generation valve (82%), heavy calcification of the native valve (65%), lack of predilatation (16%), Sievers type-1 bicuspid aortic valve (11%), and improper valve loading (5%). Infolding resulted in severe PVL causing hemodynamic instability (29%) or cardiac arrest (12%). Postdilatation was the treatment strategy in 68%, while prosthesis replacement with a new device in 23% of cases. Device success rate was 82%. Death and stroke occurred in 3% and 12% of cases. CONCLUSIONS: Prosthetic valve infolding is typically observed after resheathing of a large-size self-expanding TAVI. When infolding is timely diagnosed, prosthesis removal and replacement should be pursued. Further studies are required to precisely define predisposing factors to prevent this complication.
No takes yet. Share an insight, caveat, or question.
Beneduce et al. (2020) conducted a systematic review in Prosthetic valve infolding during transcatheter aortic valve implantation (n=34). Transcatheter aortic valve implantation (TAVI) with self-expanding valves was evaluated. Prosthetic valve infolding during TAVI was mostly associated with ≥29-mm valves (94%) and resheathing (82%), resulting in an 82% device success rate and death or stroke in 3% and 12% of cases.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: